Carbon-13 (C13) nuclear magnetic resonance (most commonly known as carbon-13 NMR spectroscopy or 13C NMR spectroscopy or sometimes simply referred to as carbon NMR) is the application of NMR spectroscopy to carbon. It is analogous to proton NMR ( NMR) and allows the identification of carbon in an organic molecule just as proton NMR identifies hydrogen atoms. 13C NMR detects only the isotope. The main carbon isotope, does not produce an NMR signal. Although circa 1 million times less sensitive than 1H NMR spectroscopy, 13C NMR spectroscopy is widely used for characterizing organic and organometallic compounds, primarily because 1H-decoupled 13C-NMR spectra are more simple, have a greater sensitivity to differences in the chemical structure, and, thus, are better suited for identifying molecules in complex mixtures. At the same time, such spectra lack quantitative information about the atomic ratios of different types of carbon nuclei, because the nuclear Overhauser effect used in 1H-decoupled 13C-NMR spectroscopy enhances the signals from carbon atoms with a larger number of hydrogen atoms attached to them more than from carbon atoms with a smaller number of H's, and because full relaxation of 13C nuclei is usually not attained (for the sake of reducing the experiment time), and the nuclei with shorter relaxation times produce more intense signals.
The major isotope of carbon, the 12C isotope, has a spin quantum number of zero and so is not magnetically active and therefore not detectable by NMR. 13C, with a spin quantum number of 1/2, is not abundant (1.1%), whereas other popular nuclei are 100% abundant, e.g. 1H, 19F, 31P.
On the other hand, the sensitivity of 13C NMR spectroscopy benefits to some extent from nuclear Overhauser effect, which enhances signal intensity for non-quaternary 13C atoms.
The chemical shift reference standard for 13C is the carbons in tetramethylsilane (TMS), whose chemical shift is set as 0.0 ppm at every temperature.
Colors =
id:nmrbar value:rgb(0.9,0.9,0.65) id:gray value:rgb(0.85,0.85,0.85)
Period = from:-1 till:220
PlotArea = left:60 bottom:20 top:10 right:10
DateFormat = yyyy
TimeAxis = orientation:horizontal format:yyyy order:reverse
ScaleMajor = gridcolor:gray unit:year increment:20 start:0
PlotData=
width:20 bar:Aldehydes from:180 till:220 color:nmrbar at:180 align:left text:R(CO)R shift:5,-5 at:220 align:left text:Aldehydes_and_ketones shift:5,-20 bar:Carboxylic from:160 till:185 color:nmrbar at:160 align:left text:R(CO)X shift:5,-5 at:185 align:left text:Carboxylic_acid_derivatives shift:0,-20 bar:Nitrile from:115 till:125 color:nmrbar at:125 text:"Nitrile RCN" align:right shift:-10,-5 bar:CC from:110 till:150 color:nmrbar at:150 text:"C=C" align:right shift:-10,-5 bar:Alkyne from:65 till:90 color:nmrbar at:90 text:"Alkyne R-CC-R" align:right shift:-10,-5 bar:RCH2O from:50 till:90 color:nmrbar at:90 text:"R-CH2-O" align:right shift:-10,-5 bar:R4C from:30 till:45 color:nmrbar at:45 text:"R4C" align:right shift:-10,-5 bar:R3CH from:30 till:50 color:nmrbar at:50 text:"R3CH" align:right shift:-10,-5 bar:RCH2X from:20 till:50 color:nmrbar at:50 text:"R-CH2-X~X= C=C, C=O, Br, Cl, N" align:right bar:R2CH2 from:20 till:30 color:nmrbar at:30 text:"R2CH2" align:right shift:-10,-5 bar:RCH3 from:5 till:20 color:nmrbar at:20 text:"RCH3" align:right shift:-10,-5 bar:TMS from:1 till:-1 color:nmrbar at:0 text:"TMS" align:right shift:-7,-2
Typical chemical shifts in 13C-NMR
The trends in J(1H,13C) and J(13C,13C) are similar, except that J(1H,13C are smaller owing to the modest value of the 13C nuclear magnetic moment. Values for 1J(1H,13C) range from 125 to 250 Hz. Values for 2J(1H,13C) are near 5 Hz and often are negative.
Some NMR probes, called cryoprobes, offer 20x signal enhancement relative to ordinary NMR probes. In cryoprobes, the RF generating and receiving electronics are maintained at ~ 25K using helium gas, which greatly enhances their sensitivity. The trade-off is that cryoprobes are costly.
In further contrast to 1H NMR, the intensities of the signals are often not proportional to the number of equivalent 13C atoms. Instead, signal intensity is strongly influenced by (and proportional to) the number of surrounding spins (typically 1H). Integrations are more quantitative if the delay times are long, i.e. if the delay times greatly exceed relaxation times.
The most common modes of recording 13C spectra are proton-noise decoupling (also known as noise-, proton-, or broadband- decoupling), off-resonance decoupling, and gated decoupling. These modes are meant to address the large J values for 13C - H (110–320 Hz), 13C - C - H (5–60 Hz), and 13C - C - C - H (5–25 Hz) which otherwise make completely proton coupled 13C spectra difficult to interpret.
With proton-noise decoupling, in which most spectra are run, a noise decoupler strongly irradiates the sample with a broad (approximately 1000 Hz) range of radio frequencies covering the range (such as 100 MHz for a 23,486 gauss field) at which protons change their nuclear spin. The rapid changes in proton spin create an effective heteronuclear decoupling, increasing carbon signal strength on account of the nuclear Overhauser effect (NOE) and simplifying the spectrum so that each non-equivalent carbon produces a singlet peak.
Both the atoms, carbon and hydrogen exhibit spins and are NMR active. The nuclear Overhauser Effect is in general, showing up when one of two different types of atoms is irradiated while the NMR spectrum of the other type is determined. If the absorption intensities of the observed (i.e., non-irradiated) atom change, enhancement occurs. The effect can be either positive or negative, depending on which atom types are involved.
The relative intensities are unreliable because some carbons have a larger spin-lattice relaxation time and others have weaker NOE enhancement.
In gated decoupling, the noise decoupler is gated on early in the free induction delay but gated off for the pulse delay. This largely prevents NOE enhancement, allowing the strength of individual 13C peaks to be meaningfully compared by integration, at a cost of half to two-thirds of the overall sensitivity.
With off-resonance decoupling, the noise decoupler irradiates the sample at 1000–2000 Hz upfield or 2000–3000 Hz downfield of the proton resonance frequency. This retains couplings between protons immediately adjacent to 13C atoms but most often removes the others, allowing narrow multiplets to be visualized with one extra peak per bound proton (unless bound methylene protons are non-equivalent, in which case a pair of doublets may be observed).
is an NMR method used for determining the presence of primary, secondary and tertiary carbon atoms. The DEPT experiment differentiates between CH, CH2 and CH3 groups by variation of the selection angle parameter (the tip angle of the final 1H pulse): 135° angle gives all CH and CH3 in a phase opposite to CH2; 90° angle gives only CH groups, the others being suppressed; 45° angle gives all carbons with attached protons (regardless of number) in phase.
Signals from quaternary carbons and other carbons with no attached protons are always absent (due to the lack of attached protons).
The polarization transfer from 1H to 13C has the secondary advantage of increasing the sensitivity over the normal 13C spectrum (which has a modest enhancement from the nuclear overhauser effect (NOE) due to the 1H decoupling).
Even though this technique does not distinguish fully between CHn groups, it is so easy and reliable that it is frequently employed as a first attempt to assign peaks in the spectrum and elucidate the structure. Additionally, signals from quaternary carbons and other carbons with no attached protons are still detectable, so in many cases an additional conventional 13C spectrum is not required, which is an advantage over DEPT. It is, however, sometimes possible that a CH and CH2 signal have coincidentally equivalent chemical shifts resulting in annulment in the APT spectrum due to the opposite phases. For this reason the conventional 13C{1H} spectrum or HSQC are occasionally also acquired.
|
|